Step Action
CAUTION
H
o
t surface. Avoid direct contact with the heating
block and the top surface of the rotating block during or
after heating. Contact with hot surfaces during or
directly after a procedure in which the heaters are on
can cause burns.
7 Tap anywhere on the screen outside the He
ater Control dialog to exit the
heater control.
8 Wait until the temperature of the heater or heaters displayed in the bottom
ro
w of the heating control indicator reaches the setpoint displayed in the top
row.
Note:
Preheating generally takes less than ten minutes.
Proceed to the next section to start a run.
Tip:
Preheat the reagents in a wet or dry bath while you wait for the syringe
he
aters to reach the setpoint temperature. Make sure to heat the empty or
filled syringes.
6 Operation
6.3 Run a procedure
6.3.3 Optional: Preheat the system
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 86