Step Action
4 Insert the R, C, and optional L syringes into the respective channels of the
rotating block.
5 Secure the syringes to the cartridge by doing one of the following:
a. Syringes with Luer fittings: Twist the syringe to engage the Luer connec-
tion.
NOTICE
Do
not tighten the Luer connection too much. Tight-
ening the connection too much can damage the
syringe and the cartridge.
b. Syringes without Luer fittings: Push the syringe until you meet resist-
an
ce.
NOTICE
Do
not insert the syringe too far. Inserting the syringe
too far can damage the syringe and the cartridge.
6 Operation
6.
3 Run a procedure
6.3.6 Load syringes and collection tubes
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 95