124 | SEVEN PLUS Continuous Glucose Monitoring System
PARAMETER PERFORMANCE CHARACTERISTICS
Frequency Allocation MICS Band
TX/RX Frequency 402.142 MHz
Bandwidth 300 kHz
Maximum Output Power 25 uW EIRP
Modulation On-Off Key
Data Rate 8192 bits/Sec
Total Packet 76 bits
Transmit Duty Cycle 9.28 ms every 5 minutes
Data Detection Range 5 ft
Average Number of Collisions due to
Coexistence. 20 DexCom Transmitters
within 15 feet over a 24 hr period
(288 transmissions each)
Total of 4 collisions, if collision occurs, no glucose
value is received
10