SEVEN PLUS User’s Guide | 125
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Transmitter is intended for use in the electromagnetic environment specied below. The cus-
tomer or the user of the Transmitter should assure that it is used in such an environment
Immunity Test IEC 60601
Test Level
Transmitter
Compliance
Level
Electromagnetic
Environment
Guidance
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV
Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30 %
Electrical Fast
Transient /burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not
Applicable-
Battery
Operated
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
Not
Applicable-
Battery
Operated
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply Input
Lines
IEC 61000-4-11
< 5 % UT (>95 % dip in UT) for
0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT) for
25 cycles
< 5% UT (>95% dip in UT)
for 5 sec
Not
Applicable-
Battery
Operated
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power Frequency
magnetic elds should be
at levels characteristic of a
typical location in a typical
commercial or hospital envi-
ronment