English
8 Supplement Scio Four modules
1 Connect the power supply with the "X16" connector to the rear
of the gas analyzer.
2 Connect the power cable to the rear of the power supply.
3 Plug the power cable into the mains power socket on the wall.
Establishing potential equalization
Differences in electrical potential between devices can be reduced
by potential equalization.
Potential equalization does not replace the protective ground con-
nection. During operation, the potential equalization connectors
must be readily accessible and the connection must be able to be
disconnected without the use of tools.
Connecting the potential equalization cable
1 Connect the potential equalization cable to the potential equal-
ization pin on the device (see page 6).
2 Connect the potential equalization cable to a potential equal-
ization connector of the hospital (e.g., wall, ceiling supply unit).
Connecting the gas analyzer to a patient monitor
The gas analyzer connects to the following patient monitors:
– Infinity Acute Care System
– Infinity Delta Series
– Infinity Gamma XL
– Infinity Gamma XXL
– Infinity Kappa XLT
(only Scio Four plus and Scio Four Oxi plus)
– Infinity Vista XL
– Vista 120
1 Connect and secure data interface cable with the "RS232C
X2" connector (A) on the back panel of the gas analyzer.
2 Connect the data interface cable with the connector at the
Infinity Docking Station or directly with the patient monitor.
Water trap
Empty or exchange the water trap in accordance with its instruc-
tions for use.
003
NOTE
Unplug the power cable from the mains power socket to discon-
nect the gas analyzer from power supply (e. g. for maintenance
or emergency reasons).
002
WARNING
Risk of infection
The liquid in the water trap may be contaminated and must
be handled and disposed of with care.
Dispose of the liquid in an adequate way and in compliance
with local regulations.
WARNING
Risk of infection
If the sample line remains connected while the water trap is
removed, contaminated liquid may escape from the water
trap.
Disconnect sample line before removing the water trap
from the medical device.
WARNING
Risk due to inaccurate gas measurement values and gas
measurement failure
Silicon can get into the measuring cuvette and influence the
gas measurement permanently.
Do not spray the O-rings of the water trap holder with sili-
con spray.
CAUTION
Risk of device failure
Do not operate the medical device without water trap.
CAUTION
Risk due to gas measurement failure
When the water trap is full, gas measurement can fail.
Check the level indicator in the water trap regularly and empty it
as necessary.
CAUTION
Risk due to gas measurement failure
Contaminants and damage of the water trap can influence gas
measurement.
Check and observe the replacement intervals of the water trap
and replace it as necessary.