Technical data
Operation characteristic values
Dimensions with water trap
Height 115 mm (4.8 in)
Width 190 mm (7.5 in)
Depth 270 mm (10.5 in)
Weight
Scio Four 2.9 kg (6.4 lbs)
Scio Four Oxi 3.2 kg (7.0 lbs)
Scio Four plus 2.9 kg (6.4 lbs)
Scio Four Oxi plus 3.2 kg (7.0 lbs)
Mode of operation Continuous
Device noise emissions 45 db(A)
Cooling Fan, temperature-regulated
Mains power supply 100 to 240 V~ 50/60 Hz
Inrush current 15-55 A
Power consumption 50 W
Housing leakage current 300 µA (per UL 60601-1:2003)
500 µA (per UL 60601-1:1988 incl. corrections)
Protection classes
Device Class 1
Electric shock protection Type BF
Penetration of liquids IP21
Classification according to MDD 93/42/EEC -
Annex IX
IIa
UMDNS Code Universal Medical Device
Nomenclature System - nomenclature system for
medical devices
17-445
Use of latex No latex (natural rubber) is used in parts with indirect or direct contact to patients
or operators.
Ambient conditions
During operation
Temperature 10 to 40 °C (50 to 104 °F)
Atmospheric pressure 700 to 1060 hPa (10.2 to 15.4 psi), with automatic barometric pressure compen-
sation
Relative humidity 20 to 95 %, without condensation
CO
2 concentration <1000 ppm
During storage and transport
Temperature –20 to +70 °C (–4 to 158 °F)
–40 to +70 °C (–40 to 158 °F) for power supply (MS18508)
Atmospheric pressure 500 to 1060 hPa (7.3 to 15.4 psi)
Relative humidity 5 to 95 %
20 to 95 %, without condensation, for power supply (MS18508)
Device outlets
Serial interfaces RS232 and MEDIBUS
Only connect devices that meet the requirements of IEC 60950-1 for ungrounded
SELV circuits or the requirements of IEC 60601-1 (as of the 2nd edition) for touch-
able secondary circuits with maximum 24 VDC nominal voltage.
Protocol STORM / MEDIBUS