Edit Menu WinAQMS Mini DAS User manual 1.0
24
Alarm Number: refers to the alarm that is triggered when the channel enters a high
or low alarm condition. One alarm number can be used for many channels.
Valid Data High and Valid Data Low: show the values at which WinAQMS Mini
DAS will label the status of the data as being bad. Setting both values to 0 will
disable the valid data limits. Data that is labelled as bad will not be included with
other valid readings in a data average labelled as „Good‟.
Channel Type: Each channel has a Channel Type, which can change depending on
the type of analyser. For example the Ecotech EC/ML 9841 nitrous oxide
analysers respond with data for NO, NO
X
and NO
2
when a data request is made
but the channel can only store one of the three values passed to it. Specifying one
of NO, NO
X
or NO
2
tells WinAQMS Mini DAS which part of the data returned
that you want. The Voltage type is the default value and is used if there is no need
to signal to the analyser that WinAQMS Mini DAS wants data specific to a
particular channel type, register etc.