WinAQMS Mini DAS User manual 1.0 Edit Menu
25
Process: The Process field determines what mathematical property of the data
received over the average period will be stored. Options include: Average
(default), Max (maximum value), Min (minimum value), Total (the sum of all the
readings received on that channel) and Std Dev (standard deviation).
Description: is a fifty-character text field. This field is used to more fully describe
a channel if the Label is insufficient
Percentage Ok: the percentage of data that must be valid before WinAQMS Mini
DAS will mark a report as being OK. If a value of 60 is entered here then if there
is a problem with over 40% of the data readings the report will have a bad status.
Log Report 1, 2, 3, 4: determines whether the channel data will be recorded in the
database for the selected report. If not ticked, the channel will still request data but
there will be Null data recorded in the database. If all four of the Log Report
options are not ticked for a channel then that channel will have no entry at all in
the database.
Out of Service: If a channel is marked as being Out of Service then the channel‟s
data will still appear in the database but its‟ status will indicate that the channel is
out of service.
Serial Port Output: specifies whether data for the channel will be included in the
string of data averages which is sent to the serial port specified by Report Settings.
Alternate File Output: specifies whether data for the channel will be included in
the data averages stored in alternative data files.
Different Analyser types will have different fields available in the Channel
Information tab. The differences are listed below.
Ecotech EC/ML 98XX, 9841, 9842, API, Dani, Environment SA and Teco
analysers do not show the Analogue Input # field.
The Digital Input Settings and Digital Output Settings tabs will only appear for
the Ecotech Termination Panel (TP) and Ecotech EC/ML analysers.
Opsis analysers have a Gas# and a Path # field instead of an Analogue Input #
field.