WinAQMS Mini DAS User manual 1.0 Appendix
65
We will now add a barometric pressure, wind speed and wind direction channel to the
channels on the TP. Switch to analyser number 1 on the Analyser Parameters
dropdown list and choose the information tab. Click on the Add Channel button three
times. Three new channels will be added to the list and they will have default values
entered into most of their columns. Make sure the settings match those of table 6
below
Table 6 Settings for the pressure, wind speed and wind direction channels
TP Pressure
TP WS
TP WD
System Channel #
3
4
5
Label
Pressure
WS
WD
Analogue Input #
2
13
3
Multiplier
60
1
72
Offset
800
0
0
Units
Mbar
m/s
deg
Decimal Places
0
1
0
High Alarm
1030
20
370
Low Alarm
980
0
-10
Alarm Number
2
3
4
Valid Data High
0
0
0
Valid Data Low
0
0
0
Channel Type
Voltage
Frequency
Voltage
Process
Average
Average
Average
Description
Barometric Pressure
Wind Speed
Wind Direction
Log Report 1
Set (ticked)
Set (ticked)
Set (ticked)
Log Report 2
Set (ticked)
Set (ticked)
Set (ticked)
Log Report 3
Set (ticked)
Set (ticked)
Set (ticked)
Log Report 4
Set (ticked)
Set (ticked)
Set (ticked)
Out of Service
Unset (not ticked)
Unset (not ticked)
Unset (not ticked)
Serial Port Output
Set (ticked)
Set (ticked)
Set (ticked)
Alternate File Output
Set (ticked)
Set (ticked)
Set (ticked)
Now press the OK button on the Analyser Parameters window. This will save the
settings to the database. You may notice that the Restart Server button is not enabled
at this time. This allows WinAQMS Mini DAS to have sufficient time to write the
settings to the database before you restart the server again with the updated settings.
The wind speed and direction sensor that is being used is an analogue output MetOne-
034 sensor.