WinAQMS Mini DAS User manual 1.0 Appendix
69
Add another new analyser to be the 9850 and changes its settings to match table 10
and a new channel to the analyser to match table 11.
Table 10 Settings for the Monitor Labs 9850 Sulfur Dioxide analyser.
Ecotech EC 9850
INFORMATION
Label
9850
Manufacturer
Ecotech
Model
98XX
Device ID
50
Data Request Period
10
Max Response Time
1
Request Delay
0
COMMUNICATION SETTINGS
Connection Type
Serial RS232
Serial Port
3
Baud Rate
9600
Data Bits
8
Parity
None
Stop Bits
1
Table 11 Settings for the Sulfur Dioxide channel
9850 SO
2
Channel
System Channel #
11
Label
SO
2
Multiplier
1
Offset
0
Units
ppm
Decimal Places
3
High Alarm
0.1
Low Alarm
0
Alarm Number
11
Valid Data High
0
Valid Data Low
0
Process
Average
Description
Sulfur Dioxide
Log Report 1
Set (ticked)
Log Report 2
Set (ticked)
Log Report 3
Set (ticked)
Log Report 4
Set (ticked)
Out of Service
Unset (not ticked)
Serial Port Output
Set (ticked)
Alternate File Output
Set (ticked)
We have eleven channels on the system, each of which will log the appropriate
analyser on the appropriate communications port. Now we will add two digital output
channels that can be used to switch solenoids when the SO
2
analyser goes into
calibration.