Patient Monitor User Manual Monitoring ECG/ RESP
- 113 -
ARR ANALYSE Menu
Figure 13-13 ARR ANALYSE
ARR ANALYSE: Pick ON during monitoring. It is set to OFF by default.
PVCs ALM: Pick ON to enable prompt message when alarm occurs; pick OFF to disable
the alarm function, and there will be a beside PVCs.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
ALM LEV: Selectable from HIGH, MED, LOW. Level HIGH represents the most serious
case.
ALM REC: pick ON to enable report printing upon PVCs alarm.
PVCs alarm is activated when the PVCs exceeds the set PVCs ALM HI value.
ARR RELEARN Pick this item to start a learning procedure.
ARR ALARM Pick this item to access the ARR ALARM dialog box to set arrhythmia
alarm parameters.
You can pick ALL ALM ON to enable the alarm function of all arrhythmia types and pick
ALL ALM OFF to disable this function. Likewise, you can pick ALL REC ON to enable the
recording function of all arrhythmia types and pick ALL REC OFF to disable this function.
Changing the ALM LEV can reset the alarm level of all arrhythmia types to the same value.
ARR RECALL Pick this item to review and edit the ARR analysis result.
The latest arrhythmia events (up to 60) are displayed.