Patient Monitor User Manual System Menu
- 35 -
4.10 Alarm Volume
The system provides 10 levels of alarm volume and audio alarm off function. The system will
give audio alarm prompt (except alarm sound) based on the selection.
The user may select different levels of volume as per clinical requirement. The method is listed
below:
Press ALARM SETUP item in SYSTEM SETUP menu to call up ALARM SETUP sub-menu
as shown below, in which the user may set up the alarm volume and other alarm information.
- ALARM VOL: Set the alarm volume by turning the knob. The valid range is from 1 to 10.
- ALM REC TIME: Set to 8 s, 16 s or 32 s.
- ALM Reset: Reset the alarm by selecting this soft key.
You can also set alarm parameters in MAINTAIN > USER MAINTAIN > ALARM SETUP.
Refer to Chapter6 Alarm for details.
4.11 Key Volume
Select SELECTION item in SYSTEM MENU to call up SELECTION sub-menu as shown
below. Select KEY VOL item and set the volume. The selections are OFF, LOW, MED,
HIGH.
Figure 4-12 Selection
4.12 Drug Calculation
The monitor provides drug calculation and titration table display functions for fifteen different
drugs. For details, please refer to the Chapter11 Drug Calculation and Titration Table.
4.13 Waveform Demonstration
Select DEMO item in SYSTEM MENU to call up INPUT DEMO KEY. After entering the
password, the system enters the Demonstration Waveform status.
The purpose of waveform demonstration is only to demonstrate the machine performance and for
training purposes. In clinical applications, this function is not recommended because the DEMO
will mislead the hospital personnel to treat the waveform and parameter as actual data of the
patient, which may result in delay of treatment or mistreatment.