Patient Monitor User Manual Alarm Information
- 66 -
Message
Cause
Alarm level
What to do
ECG V LEAD
OFF
ECG electrode V falls
off the skin or the ECG
cable V falls off the
monitor.
Low
ECG SIGNAL
EXCEED
ECG measuring signal
is beyond measuring
range.
Low
Check lead connection and
patient condition
ECG COMM
STOP
ECG module failure or
communication failure
High
Stop using measuring
function of ECG module;
notify biomedical engineer or
manufacturers service staff.
ECG NOISE
ECG measuring signal
is greatly interrupted.
Low
Check lead connection and
patient condition
7.3 Prompts
Prompt message:
Message
Cause
Manual measuring…
It is in the process of measuring.
Continual Measuring
In continuous measuring mode.
Auto measuring…
It is in the process of automatic measuring.
Measurement over
Measurement is over.
Calibrating…
It is in the process of calibrating.
Calibration over
Calibration is over.
Pneum testing…
It is in the process of pneumatic testing.
Pneum test over
Pneumatic test is over.
Resetting...
NIBP module is resetting.
Please Start
NIBP module is in idle status.
CO
2
STANDBY
STATUS
Turn from measuring mode to standby mode,
making the module in energy-saving status.
CO
2
SENSOR
WARMUP
The CO
2
module is at warm-up state