Patient Monitor User Manual EMC Information
- 183 -
Appendix II EMC Information
- Guidance and Manufacture’s Declaration
A2.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
The patient monitors of iM8 & M8 series patient monitor are intended for use in the
electromagnetic environment specified below. The customer or the user of the iM8 & M8
series patient monitor should assure that they are used in such an environment.
Electromagnetic environment – guidance
The patient monitors of iM8 & M8 series use RF
energy only for internal function. Therefore, RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The patient monitors of iM8 & M8 series are
suitable for use in all establishments, other than
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
NOTE:
The EMISSIONS characteristics of iM8 & M8 series make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) iM8 & M8 series might not offer adequate
protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
A2.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The patient monitors of iM8 & M8 series are intended for use in the electromagnetic
environment specified below. The customer or the user of the iM8 & M8 series patient monitors
should assure that they are used in such an environment.