Patient Monitor User Manual EMC Information
- 185 -
A2.3 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The patient monitors of iM8 & M8 series are intended for use in the electromagnetic
environment specified below. The customer or the user of iM8 & M8 series patient monitors
should assure that they are used in such an environment.
Electromagnetic environment -
guidance
Conducted RF
IEC/EN
61000-4-6
Radiated RF
IEC/EN
61000-4-3
3 V
rms
150 kHz to 80 MHz
6Vrms
c
in ISM
bands between
0,15 MHz and 80
MHz
3 V/m
80 MHz to 2.7 GHz
See Table 1
3 V
rms
150 kHz to
80 MHz
6Vrms
c
in
ISM bands
between
0,15 MHz
and 80 MHz
3 V/m
80 MHz to
2.7 GHz
Comply with
Table 1
Portable and mobile RF
communications equipment should
be used no closer to any part of the
iM8 & M8 series patient monitors,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals
such as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the monitor, including cables
specified by the manufacturer).
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the