- 147 -
18.2.1 Zeroing the sensor
You must perform zeroing following the steps when using the new airway adapter.
1. Expose the sensor to room air and keep it away from all sources of CO
2
including the
ventilator, the patient’s breath and the operator’s.
2. In the CO
2
SETUP menu, please set the WORK MODE to MEASURE.
3. For EDAN module, select User Maintain > CO
2
Maintain, and click Zero, or select
ZERO CAL. in the CO
2
SETUP menu. For Respironics modules, click ZERO CAL. in
the CO
2
SETUP menu.
4. The messages indicate: “zeroing”, “zero successfully”. After the zero calibration is
finished, the user can start CO
2
Monitoring.
NOTE:
For EDAN module, automatic zero is performed when turning on the monitor.
18.2.2 Sidestream CO
2
Module
18.2.2.1 Measurement Steps
EDAN EtCO
2
module
1 Fix the water trap to the water trap holder on the right side of the monitor, confirm it is well
fixed.
2 Connect the sampling cannula or the sampling line to the water trap.
3 Set WORK MODE to MEASURE.
4 For intubated patients, an airway adapter is required. For non-intubated patients, place the
nasal cannula or the sampling mask onto the patient.
CAUTION
1 The water trap collects water drops condensed in the sampling line and therefore
prevents them from entering the module. If the water trap is nearly filled, you should
replace it to avoid blocking the airway.
2 Based on a sample gas temperature of 37 C, a room temperature of 23 C and
sample relative humidity of 100%, the water trap will be filled after approximately 90
hours with the flowrate of 100 ml/min and approximately 130 hours with the flowrate
of 70 ml/min. In clinical practice, the water trap can be used for a longer time before it
is filled. It is recommended to replace the water trap once every month.
3 When replacing the water trap or suspecting the measurement value, please check if
the O-rings of the water trap holder are normal and well installed. If the O-rings get
damaged or loose, contact service staff of the manufacturer.
4 To prevent the module from abnormal work, please ensure the water trap detection
button is not mistakenly touched.
5 Please replace and discard the water trap when blocking. Don’t reuse it, otherwise
the reading is not accurate and even the device may be damaged.