4.15 Data Storing
Users can store the measured data into USB storage by Data store function, query or delete data
in the menu.
The factory default of data store is in off status. If DATA STORE in FACTORY MAINTAIN
is set to ON, user can select DATA STORE in SYSTEM MENU to call up the following dialog
box:
NOTE:
The monitor may not be compatible with all models of USB disks, it is recommended to
use Netac USB disk of 4G.
Figure 4-17 Data Store Menu
DATA STORE: set this item to ON or OFF to turn on or off the data store function.
The data file will be stored into the folder of patient-data/patient ID in the USB storage; if the
patient ID has not been set, the data will be stored into the default folder “patient” in USB
storage.
Each data file is named by time, it can save 96-hour trend data with 1-min resolution, 1-hour
trend data with 1-second resolution, 60 groups parameter alarm, 60 groups ARR data, 500 groups
NIBP data, 720 seconds waveforms and patient information.
EXIT U DISK: users should exit the U disk via the menu before dismounting it.
After selecting EXIT U DISK, if the data is being stored, it will indicate “Transmitting…,
Please Waiting”; if the U disk is dismounted successfully, it will indicate EXIT U DISK
SUCCESS. After the USB icon vanishes, remove the U disk.