Patient Monitor User Manual Default Settings
- 192 -
A3.7 NIBP
NIBP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit (SYS)
160
120
90
Alarm Low Limit (SYS)
90
70
40
Alarm High Limit (Map)
110
90
70
Alarm Low Limit (Map)
60
50
25
Alarm High Limit (Dia)
90
70
60
Alarm Low Limit (Dia)
50
40
20
Inflation value
160
140
100
Unit
mmHg
Interval
Manual
A3.8 TEMP
TEMP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit (T1)
39.0
39.0
39.0
Alarm Low Limit (T1)
36.0
36.0
36.0
Alarm High Limit (T2)
39.0
39.0
39.0
Alarm Low Limit (T2)
36.0
36.0
36.0
Alarm High Limit (TD)
2.0
2.0
2.0
Unit
A3.9 IBP
IBP Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium

Table of Contents