Patient Monitor User Manual Alarm Information
- 65 -
Message
Cause
Alarm level
What to do
CO
2
ZERO
REQUIRED
Zero calibration failure
Low
Stop using
CO
2
alarm
function,
notify
biomedical
engineer or
Manufacturer’s
service staff.
Only
applicable
to
Respironics
CO
2
Module
CO
2
INIT ERR
CO
2
module failure
High
Stop using
CO
2
alarm
function;
notify
biomedical
engineer or
Manufacturer’s
service staff.
CO
2
SYSTEM
ROM ERR
CO
2
module failure
High
CO
2
INT RAM
ERR
CO
2
module failure
High
NetWork
Disconnect
The monitor and CMS
network is
disconnected
Low
Please check the network and
reconnect it.
ECG LEAD OFF
1) The drive electrode
or more than one ECG
limb electrode falls off
the skin;
2) ECG cables fall off
the monitor.
Low
Make sure that all electrodes,
leads and patient cables are
properly connected.
ECG LL LEAD
OFF
ECG electrode LL falls
off the skin or the ECG
cable LL falls off the
monitor.
Low
ECG LA LEAD
OFF
ECG electrode LA
falls off the skin or the
ECG cable LA falls off
the monitor.
Low
ECG RA LEAD
OFF
ECG electrode RA
falls off the skin or the
ECG cable RA falls off
the monitor.
Low