Patient Monitor User Manual Alarm Information
- 64 -
Message
Cause
Alarm level
What to do
CHECK
ADAPTER
1) For the Respironics
CO
2
module: The
cannula is off or
disconnected.
2) For the EDAN
EtCO
2
module: The
water trap is
disconnected or not
properly connected.
Low
1) For the
Respironics
CO
2
module:
Check whether
the adapter is
properly
connected or
replace the
adapter.
2) For the
EDAN EtCO
2
module:
Properly
connect the
water trap.
CO
2
ADAPTER
OCCLUDE
1) For the Respironics
CO
2
module: The
cannula/gas outlet is
occluded.
2) For the EDAN
EtCO
2
module: The
water trap/ cannula/gas
outlet is occluded.
High
1) For the
Respironics
CO
2
module:
Replace the
cannula;
Check whether
the gas outlet
is occluded.
2) For the
EDAN EtCO
2
module:
Replace the
water trap;
Check whether
the cannula is
curved, or
whether the
gas outlet is
occluded.