Patient Monitor User Manual Alarm Information
- 63 -
Message
Cause
Alarm level
What to do
notify biomedical engineer or
Manufacturers service staff.
IBP Catheter Off
IBP catheter falls off
due to patient
movement.
High
Check the catheter
connection and reconnect it.
IBP Sensor Err
Malfunction in IBP
sensor or in the
extension cable.
Medium
Replace the IBP sensor or
the extension cable.
CO
2
SENSOR
OFF
CO
2
sensor falls off
Low
Well connect the sensor
CO
2
WATERTRAP
OCCLUDE
Water trap of
SideStream is occluded
High
Make sure the gas exhaust
works well
CO
2
COMM
STOP
CO
2
module failure or
communication failure
High
Stop using
CO
2
alarm
function,
notify
biomedical
engineer or
Manufacturer’s
service staff.
Applicable
to EDAN
EtCO
2
module and
Respironics
CO
2
Module
CO
2
Sensor Over
Temp
CO
2
module exceeds
the working
temperature range.
High
Stop using
measuring
function of
CO
2
module,
notify
biomedical
engineer.
CO
2
SENSOR
FAULT
CO
2
module failure
High
Stop using
measuring
function of
CO
2
module,
notify
biomedical
engineer.