Patient Monitor User Manual Monitoring TEMP
- 135 -
Figure 16-1 TEMP SETUP
ALM: pick ON to enable prompt message during the TEMP alarm; pick OFF to disable the
alarm function, and prompt the symbol besides TEMP numeric.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
ALM LEV: used to set up the alarm level, selectable from HIGH, MED or LOW.
ALM REC: used to start/stop recording TEMP alarms. Pick ON to enable report printing
upon TEMP alarm.
Alarm for T1, T2, TD occurs when the measured temperature exceeds the set alarm upper
limit or falls below alarm lower limit.
T1 represents the TEMP of Channel 1, T2 represents the TEMP of Channel 2, TD represents the
TEMP difference of T1 and T2.
UNIT: To set temperature unit (ºC or ºF).
DEFAULT: Pick this item to access the TEMP DEFAULT CONFIG dialog box, in which
the user may select whether the FACTORY DEFAULT CONFIG or the USER
DEFAULT CONFIG is to be used. After selecting any of the items and exiting the dialog
box, the system will pop up the dialog box asking for the user’s confirmation.