Patient Monitor User Manual Intended Use and Safety Guidance
- 6 -
WARNING
43 To protect the monitor from damage during defibrillation, for accurate measurement
information and to protect against noise and other interference, use only accessories
specified by the manufacturer.
44 If several items of medical equipment are interconnected, pay attention to the sum of
the leakage currents, otherwise it may cause shock hazard. Consult your service
personnel.
45 No modification of this equipment is allowed without authorization of the
manufacturer. If this equipment is modified, appropriate inspection and testing must
be conducted to ensure continued safe operation.
46 The device must be connected to the ground to avoid signal interference.
47 The monitor is suitable for use in the presence of electrosurgery. When the monitor is
used with HF surgical equipment, user (doctor or nurse) should be cautious about
patient safety.
48 Make sure networking function is used in a secure network environment.
49 Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the monitor, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
CAUTION
1 Federal laws (U.S.) restrict this device to sale, distribution and use by, or on the order
of a physician.
2 Electromagnetic Interference - Ensure that the environment in which the patient
monitor is installed is not subject to any sources of strong electromagnetic
interference, such as radio transmitters, mobile telephones, microwaves, etc.
3 Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine,
dust area, high temperature and humid environment.
4 Do not immerse transducers in liquid. When using solutions, use sterile wipes to
avoid pouring fluids directly on the transducer.
5 Do not use autoclave or gas to sterilize the monitor, recorder or any accessories.
6 The device and reusable accessories may be sent back to the manufacturer for
recycling or proper disposal after their useful lives.
7 Remove a battery whose life cycle has expired from the monitor immediately.
8 Assembly of the monitor and modifications during actual service life shall be
evaluated based on the requirements of IEC60601-1.