Patient Monitor User Manual Monitoring IBP (Optional)
- 142 -
Figure 17-6 IBP Calibration Menu
Calibrate the transducer:
Turn the knob to select the item CH1 CAL VALUE, press and turn the knob to select the
pressure value to be calibrated for channel 1. Then turn the knob to select CALIBRATE in the
menu to start calibrating channel 1.
Turn the knob to select the item CH2 CAL VALUE, press and turn the knob to select the
pressure value to be calibrated for channel 2. Then turn the knob to select CALIBRATE in the
menu to start calibrating channel 2.
The pressure calibration of the portable patient monitor
1: Hydrargyrum pressure meter; 2: 3-way connector; 3: 3-way stopcock; 4: Pressure transducer; 5:
Pressure transducer interface cable; 6: Monitor
Figure 17-7 IBP Calibration
6
5
4
SET UP
FREEZE
SILENCE
RECORD
START
MENU
POWER
ALARM
CHARGE
3
2
1