Patient Monitor User Manual Monitoring IBP (Optional)
- 144 -
“IN DEMO, FAIL!”
Make sure that the monitor is not in DEMO mode. Contact service technician if
necessary.
“PRESSURE OVER RANGE, FAIL!”
Make sure that you have selected transducer value in IBP CAL, then start the calibration.
IBP SCALE ADJUST Submenu
Figure 17-8 IBP SCALE ADJUST Menu
The waveform and corresponding scale appears in the IBP Waveform Area with 3 dotted lines
representing High Limit Scale, Reference Scale, and Low Limit Scale from the top to the bottom.
Values of the three scales can be user-set according to the instruction given below.
IBP label: selectable from ART, PA, CVP, RAP, LAP, ICP, P1, P2;
HI: IBP value of High Limit scale, its range is the measuring range of the current pressure.
NOTE:
The HI value must be higher than the LO value.
LO: IBP value of Low Limit scale, its range is the measuring range of the current pressure.
NOTE:
The LO value must be lower than the HI value.
VAL: IBP value of Reference scale (between HI and LO).
NOTE:
When change HI scale, Low scale or Reference scale of IBP waveform and the
corresponding IBP waveforms are displayed under the menu window, the waveform will
come penetratingly through the menu window for observing.