EasyManuals Logo

EDAN M8 Series User Manual

EDAN M8 Series
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
Patient Monitor User Manual Monitoring CO
2
(Optional)
- 149 -
Figure 18-3 Air adapter
For non-intubated patients: Place the nasal cannula onto the patient.
Figure 18-4 Place the nasal cannula
NOTE:
1 You must perform a zero calibration as described in this procedure each time the
ambient temperature changes more than 10 °C (for example during transport).
2 Always connect the airway adapter to the sensor before inserting the airway adapter
into the breathing circuit. In reverse, always remove the airway adapter from the
breathing circuit before removing the sensor.
3 Always disconnect the cannula, airway adapter or sample line from the sensor when
the sensor is not in use.
4 To extend the lifetime of the module, set Work Mode to Standby when the module is
not in use.
5 The sidestream CO
2
module continuously extracts a quantity of gas from the
patient's airway per minute. Please do not use this module in any patient who will be
affected by this sampling rate.
6 If the catheter falls off during the measurement, it is necessary to re-zero after the
catheter is well connected, and then measurement can be performed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M8 Series and is the answer not in the manual?

EDAN M8 Series Specifications

General IconGeneral
BrandEDAN
ModelM8 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals