Patient Monitor User Manual Alarm Information
- 60 -
adjust the electrode position
on the patient in order to
reduce the interference of
cardiogenic artifact.
RR cannot be
measured due to
patient movement.
Check whether the RESP
leads are well connected.
Keep the patient calm for
better monitoring.
SpO
2
sensor was not
connected well, or the
connection is loose.
Make sure the sensor is well
connected to the patient’s
finger or other parts.
SpO
2
sensor was not
connected well, or the
connection is loose.
Make sure the monitor and
sensor are well connected.
Reconnect the sensor.
SpO
2
module failure
or communication
failure
Stop using measuring
function of SpO
2
module,
notify biomedical engineer or
Manufacturer’s service staff.
The pulse signal is too
weak or the perfusion
of the measurement
site is too low.
Reconnect the SpO
2
sensor
and change the measurement
site. If problem exists, please
notify biomedical engineer or
manufacturer’s service staff.
There is interference
with SpO
2
measurement signals
and the waveform is
abnormal.
Check the condition of
patient and avoid patient
movement; make sure the
cable is well connected.
There is interference
with SpO
2
measurement signals
due to patient
movement, ambient
light, electrical
interference or else.
Reduce interference of the
ambient light and avoid
sensor’s exposure to strong
light.
Malfunction in the
SpO
2
sensor or in the
extension cable.
Replace the SpO
2
sensor or
the extension cable.
NIBP module failure
or communication
failure
Stop using measuring
function of NIBP module,