Patient Monitor User Manual Alarm Information
- 40 -
Message
Cause
Alarm Level
Action Taken
NIBP Comm Fail
NIBP module failure or
communication failure
High
Stop using
measuring function
of NIBP module, and
notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Leak
NIBP pump, valve, cuff or
tube has a leakage.
Low
Check the
connections and the
wrapped cuff to see
whether they are all
prepared well.
NIBP Excessive
Pressure
Pressure has exceeded the
specified upper safety limit.
Low
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Init Pressure
High
The initial pressure is too
high during measuring
Low
NIBP Aux Excessive
Pressure
Pressure has exceeded the
second safety limit as
specified.
High
Notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Time Out
Measuring time has exceeded
the specified time.
Low
Measure again or use
other measuring
method.
NIBP Self Test
Error
Sensor or other hardware
errors.
Low
If failure persists,
stop using measuring
function of NIBP
module and notify
biomedical engineer
or manufacturer’s
service staff.