Patient Monitor User Manual Alarm Information
- 41 -
NIBP System
Pressure
Abnormality
Atmospheric pressure or
system pressure is abnormal.
The valve is occluded so that
deflation is failed.
Check whether the
airway is occluded
or pressure sensor
works properly. If
the problem still
exists, contact your
service personnel.
Contact your service
personnel.
Cuff is not properly wrapped
or no cuff is connected.
Cuff is too loose or patient
pulse is too weak.
Use other methods to
measure blood
pressure.
All of SYS, DIA and MAP
values are beyond the
measurement range.
SYS (NIBP) value is beyond
the measurement range.
DIA (NIBP) value is beyond
the measurement range.
MAP (NIBP) value is
beyond the measurement
range.
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Make sure that the
patient under
monitoring is
motionless.
Fail to deflate normally
during the leak test, so NIBP
leak test cannot be finished.
Test again. If the
problem still exists,
contact your service
personnel.
Maybe pre-inflation pressure
can’t block the vessel.
Check if the cuff
leak or cuff is
properly wrapped.
Arrhythmia is serious or
pulse rate is not regular due
to the patient movement.
Make sure that the
patient under
monitoring is
motionless.