Patient Monitor User Manual Alarm Information
- 41 -
Message
Cause
Alarm Level
Action Taken
NIBP System
Pressure
Abnormality
Atmospheric pressure or
system pressure is abnormal.
The valve is occluded so that
deflation is failed.
Low
Check whether the
airway is occluded
or pressure sensor
works properly. If
the problem still
exists, contact your
service personnel.
NIBP System
Failure
NIBP is not calibrated.
High
Contact your service
personnel.
NIBP Loose Cuff
Cuff is not properly wrapped
or no cuff is connected.
Low
Properly wrap the
cuff.
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Low
Use other methods to
measure blood
pressure.
NIBP Range
Exceeded
All of SYS, DIA and MAP
values are beyond the
measurement range.
High
SYS(NIBP)
Overrange
SYS (NIBP) value is beyond
the measurement range.
High
DIA(NIBP)
Overrange
DIA (NIBP) value is beyond
the measurement range.
High
MAP(NIBP)
Overrange
MAP (NIBP) value is
beyond the measurement
range.
High
NIBP Interference
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Low
Make sure that the
patient under
monitoring is
motionless.
NIBP Leak Test
Error
Fail to deflate normally
during the leak test, so NIBP
leak test cannot be finished.
Low
Test again. If the
problem still exists,
contact your service
personnel.
NIBP Pressure Low
Maybe pre-inflation pressure
cant block the vessel.
Low
Check if the cuff
leak or cuff is
properly wrapped.
NIBP Pulse
Abnormal
Arrhythmia is serious or
pulse rate is not regular due
to the patient movement.
Low
Make sure that the
patient under
monitoring is
motionless.