Patient Monitor User Manual Alarm Information
- 47 -
Message
Cause
Please Press 'Zero'.
Enter the IBP zeroing menu, and zeroing is not performed
yet.
Zero OK
IBP completes zeroing.
Pulsatile Pressure Zero Fail.
During the zeroing process, pressure fluctuation is
excessive.
Pressure out of normal
range,Fail.
During the zeroing process, pressure value is beyond the
zeroing range.
Sensor Off, Fail!
Perform zeroing when the sensor is off.
Invalid Time,Zero Fail.
Time is not set up prior zeroing.
Unable to Calibrate in Demo
Mode
Perform zeroing in Demo Mode.
Zeroing...
Zeroing is in progress.
Please Press 'Calibrate'.
Enter the Calibration menu, and Calibration is not
performed yet.
Calibration OK
Calibration is completed.
Pulse Pressure Calibration
Failed
During the Calibration process, pressure fluctuation is
excessive.
Pressure out of range
During the Calibration process, pressure value is beyond
the Calibration range.
Zeroing and Calibration
Failed
Zeroing is not performed prior calibration.
Sensor Off, Fail.
Perform calibration when the sensor is off.
Invalid Time, Calibration Fail.
Time is not set up prior calibration.
Unable to Calibrate in Demo
Mode
Perform calibration in Demo Mode.
Calibrating...
Calibration is in progress.
C.O. Lack Param
Parameter is not configured for C.O. measurement.
Warm-up over
The monitor displays this message after taking the sensor
out of the bracket and warm-up is over.
Measure over
After the Predict measuring is over, the data and message
display on the interface.
Measure time out
No measuring result after the module entering Predict state
for 30 s.
Printer Busy
The monitor is performing a print job.