E-2
System Care, Service and Support
The ClearSight Finger Cuff is designed for single patient
use only. Do not attempt to clean and reuse the ClearSight
Finger Cuff on more than a single patient.
Cleaning the Patient Cables and Connectors
The patient cables contains electrical and mechanical compo-
nents and is therefore subject to normal use wear and tear.
Visually inspect the cable insulation jacket, strain relief and
connectors before each use. If any of the following conditions
are present, discontinue use of the cable.
• Broken insulation
• Frays
• Connector pins are recessed or bent.
• Connector is chipped and/or cracked.
The patient cables are not protected against fluid ingress. To
clean, wipe the cable with a damp, soft cloth using 10%
bleach and 90%water solution as needed and then air dry the
connector. Please contact Technical Support or your local
Edwards Representative for further assistance.
Service and Support
See the Troubleshooting chapter for diagnosis and remedies.
If this information does not solve the problem, contact
Edwards Lifesciences.
Edwards provides monitor operations support:
• Within the United States and Canada, call
1.800.822.9837.
• Outside the United States and Canada, contact your local
Edwards Lifesciences representative.
• E-mail operational support questions to
tech_support@edwards.com.
Have the following information before you call:
• The serial number of the Monitor, Pump-Unit and
Pressure Controller, located on the rear panel of these
units.
• Software version which is displayed at the bottom of the
screen during monitor initialization;
• The text of any error message and detailed information as
to the nature of the problem.
Figure E-1 is an example of the Startup screen.
WARNING
Do not use damaged components/sensors or
components/sensors with exposed electrical contacts to
prevent patient or user shocks.
CAUTION
Do not use any damaged system components. Use of a
damaged system component may result in inaccurate
measurements or may damage the EV1000 Clinical
Platform NI.
CAUTION
If any electrolytic solution, for example NaCl or lactated
Ringer’s solution, is introduced into the cable
connectors while they are connected to the platform,
the excitation voltage can cause electrolytic corrosion
and rapid degradation of the electrical contacts.
CAUTION
Do not immerse any cable connectors in fluid or use a
hot air gun to dry cable connectors. Refer to cleaning
instructions.
Figure E-1 Startup Screen