Monitor Display Options
6-6
Alarms / Targets Setup Screen
The Alarms / Targets Setup Screen allows you to view and set
up alarms and targets for each key parameter. The settings for
each key parameter are displayed in a parameter box. The
currently configured key parameters are the first set of key
parameters displayed. The remaining key parameters are
displayed in a defined order. The parameters also indicate
what the target ranges are based on: Custom Default,
Edwards Default, and Modified.
Visual and audible alarm settings are only applicable to
parameters being displayed.
To modify Alarms / Targets:
1 Touch the Settings button.
2 Touch Parameter Settings.
3 Touch Alarms / Targets.
4 Touch anywhere in a parameter box to display the Alarm/
Target popup for the parameter.
There is a 2 minute inactivity timer associated with this
screen.
The red, yellow and green rectangles are fixed shapes, and
don’t change size / shape.
Configure All Targets
Targets can easily be configured or changed all at the same
time. From the Configure All screen, the user can:
• Set Custom Defaults for all parameter alarm and tar-
get settings.
• Restore all parameter alarm and target settings to
Custom Defaults.
• Restore all parameter alarm and target settings to
Edwards Defaults.
• Enable or disable audible alarms for all applicable
parameters.
• Enable or disable target ranges for all parameters.
1 Touch the Settings button.
2 Touch Parameter Settings.
3 Touch Alarms / Targets.
4 Touch the Configure All button.
5 To turn audible alarms on or off for all parameters, touch
the Audible Alarm On/Off buttons and enter the
password.
6 To enable or disable all targets for parameters that support
target ranges, touch the Target On/Off buttons.
7 To restore all settings to your custom defaults, touch
Restore All to Custom Defaults. The message, “This
action will restore ALL Alarms and Targets to the
Custom Defaults.” appears.
8 Touch Continue on the confirmation popup to confirm
the restore.
9 To restore all settings to the Edwards defaults, touch
Restore All to Edwards Defaults. The message, “This
action will restore ALL Alarms and Targets to the
Edwards’ Defaults.” appears.
10 Touch Continue on the confirmation popup to confirm
the restore.
Table 6-2 Target Defaults
Default Name Description
Custom Default A custom default was set for the parameter
and the parameter has not been modified
from that default.
Edwards Default The parameter has not been changed from
the original settings.
Modified Parameter was changed for this patient.
Figure 6-8 Alarms / Targets Configuration