5-7
Navigating the EV1000 Clinical Platform NI
Cockpit Screen
This monitoring screen, shown in Figure 5-11, displays
globes with the values of the parameter being monitored.
They graphically indicate target, out of range and target
values with needle indicators to show where the patient’s
parameter falls. In addition, the value within the globe will
flash when the parameter is alarming.
The Key Parameters display a more complex target and alarm
indicator. The full display range of the parameter is used to
create a gauge from the graphical trends minimum to
maximum settings. A needle is used to indicate the current
value on the circular gauge scale. When target ranges are
enabled, red, yellow and green are used to indicate the target
and alarm regions within the circular gauge. When target
ranges are not enabled, the circular gauge area is all gray in
color and target or alarm indicators are removed. The value
indicator arrow changes to indicate when the values are out
of the gauge scale limits.
Goal Positioning Screen
The Goal Positioning Screen allows the user to monitor and
track the relationship of two key parameters by plotting them
against each other on an XY plane. See Chapter 14, "EV1000
Clinical Platform NI Advanced Features" , for more infor-
mation.
Physio Relationship
The Physio Relationship screen, as shown in Figure 5-12,
displays most of the parameters available on the system and
their relationship to each other. The screen displays lines
connecting the parameters highlighting the relationship of the
parameters to each other.
Line colors correspond to the lanterns to draw your eye to
problem areas. If a lantern turns yellow and the one above is
green, the vertical line above and the horizontal line below
turn yellow. See Chapter 8 “Physiology and Physio
Relationship Monitoring Screens” for more information.
Status Indicators
The lantern at the top of each parameter globe indicates the
patient’s current status. The color changes as the patient’s
status changes. The globes may display additional infor-
mation:
Figure 5-11 Cockpit Monitoring Screen
Figure 5-12 Physio Relationship Screen
Figure 5-13 Parameter Globe
SVV Filtering Exceeded Indicator
Audible alarm
indicator
Continuous %
Change
Indicator
Target Status
Indicator
Parameter Value
Units
Parameter Name