7-10
Methodology and Monitoring
Physiocal Control
Physiocal can be observed on the ART display as a stepwise
increase in pressure upon startup and as brief interruptions
throughout monitoring.To accurately account for changes in
the finger artery characteristics throughout monitoring,
Physiocal is performed at regular intervals resulting in
momentary interruptions to the arterial waveform. Physiocal
can be temporarily disabled. To disable Physiocal:
1 Touch the Clinical Actions button.
2 Touch More.
3 Touch Advanced Options.
4 To disable Physiocal, toggle the Physiocal button from
enabled to disabled. Physiocal will automatically be
enabled after 1 minute.
5 Toggle the Physiocal button to enabled to turn Physiocal
back on.
6 Touch the Home button.
Physiocal should not be disabled until 5 minutes have
passed from the start of monitoring. Measurements will be
most stable once the Physiocal interval status is 30 beats or
higher (‘Physiocal≥30’).