12-4
Help and Troubleshooting
Pressure Controller Communication
The Pressure Controller lights indicate the status of the Clear-
Sight Finger Cuff(s) and Heart Reference Sensor.
Battery power very low Amber Flashing ON/OFF Connect the main power cable to charge battery. If
the system is not connected to mains it will shut down
in one minute.
While connected to mains, the battery is not charged
sufficiently to provide battery operation. Keep the
system connected to mains to charge the battery.
AC POWER STATUS LIGHT
AC Mains Plugged in
Green Solid ON None
AC Mains Disconnected
No light Solid OFF None
MONITOR STATUS LIGHT
Monitor Power ON Green Solid ON None
Monitor in STANDBY Amber Solid ON Wait for the monitor to start up.
Monitor Power OFF No light Solid OFF None
Table 12-1 Pump-Unit Communication and Power Lights
Condition Color Light Pattern Suggested Action
Figure 12-8 Pressure Controller LED Indicators
Table 12-2 Pressure Controller Communication Lights
Condition Color Light Pattern Suggested Action
CUFF STATUS LIGHT
No ClearSight Finger Cuff connected No light Solid OFF None
ClearSight Finger Cuff connected Green Solid ON None. The system is ready to start a measurement.
Active monitoring Green Flashing ON/OFF None. The connected ClearSight Finger Cuff is
actively monitoring.
Defective ClearSight Finger Cuff
connected
Expired ClearSight Finger Cuff
connected
Non Edwards Finger Cuff connected
Amber Flashing ON/OFF Verify that an Edwards Finger Cuff has been used.
Disconnect and reconnect the ClearSight Finger Cuff.
Replace the finger cuff with a genuine Edwards Cuff.
Restart the measurement.
If the problem persists, contact Edwards Technical
Support.
ClearSight Finger
Cuff(s) Status
Heart Reference
Sensor Status