Help and Troubleshooting
12-7
ClearSight Faults and Alerts
Table 12-5 ClearSight Faults and Alerts
Message Possible Causes Suggested Actions
Fault: Invalid Finger Cuff #1
Connected
Non Edwards Finger Cuff #1 detected.
Defective Finger Cuff #1 connected.
Verify that an Edwards Finger Cuff has been used.
Disconnect and reconnect Edwards Finger Cuff #1.
Replace Finger Cuff #1 with a genuine Edwards Cuff.
Restart measurement.
If problem persists, contact Edwards Technical Support.
Fault: Invalid Finger Cuff #2
Connected
Non Edwards Finger Cuff #2 detected.
Defective Finger Cuff #2 connected.
Verify that an Edwards Finger Cuff has been used.
Disconnect and reconnect Edwards Finger Cuff #2.
Replace Finger Cuff #2 with a genuine Edwards Cuff.
Restart measurement.
If problem persists, contact Edwards Technical Support.
Fault: Finger Cuff Disconnected Previously connected Finger Cuff(s) not
detected.
Disconnect and reconnect Edwards Finger Cuff(s).
Replace Finger Cuff (s).
Restart measurement.
Fault: Finger Cuff #1 Error Finger Cuff #1 defective.
Poor connection between Finger Cuff #1
and Pressure Controller.
Disconnect and reconnect Edwards Finger Cuff #1.
Replace Finger Cuff #1.
Restart measurement.
If problem persists, contact Edwards Technical Support.
Fault: Finger Cuff #2 Error Finger Cuff #2 defective.
Poor connection between Finger Cuff #2
and Pressure Controller.
Disconnect and reconnect Edwards Cuff #2.
Replace Finger Cuff #2.
Restart measurement.
If problem persists, contact Edwards Technical Support.
Alert: Finger Cuff #1 Expiration
in < 5 Minutes
Finger Cuff #1 approaching maximum use
time.
Replace Finger Cuff #1 to ensure uninterrupted
measurement.
Alert: Finger Cuff #1 Has
Expired.
Finger Cuff #1 has exceeded maximum
use time.
Replace Finger Cuff #1.
Restart measurement.
Fault: Finger Cuff #1 Has
Expired. Replace Cuff
Finger Cuff #1 has exceeded maximum
use time.
Replace Finger Cuff #1.
Restart measurement.
Alert: Finger Cuff #2 Expiration
in < 5 Minutes
Finger Cuff #2 approaching maximum use
time.
Replace Finger Cuff #2 to ensure uninterrupted
measurement.
Alert: Finger Cuff #2 Has
Expired
Finger Cuff #2 has exceeded maximum
use time.
Replace Finger Cuff #2.
Restart measurement.
Fault: Finger Cuff #2 Has
Expired. Replace Cuff
Finger Cuff #2 has exceeded maximum
use time.
Replace Finger Cuff #2.
Restart measurement.
Fault: Accumulated Single Cuff
Monitoring Has Reached The
Duration Limit
Cumulative measurement on the same
finger exceeded maximum duration of 8
hours.
Remove Cuff from finger.
Place the Cuff on another finger and press 'Continue' on
the Popup.
Restart Measurement.
Fault: Check HRS Connection HRS connection not detected. Disconnect and reconnect Edwards HRS.
Replace HRS.
Alert: HRS Out of Range HRS incorrectly zeroed.
HRS measurement exceeds allowed limit.
HRS detached from Finger Cuff or
phlebostatic axis.
HRS is defective.
Rezero HRS.
Verify HRS placement -The finger end should be attached
to Finger Cuff and heart end should be placed at
phlebostatic axis.
Replace HRS.
Restart Measurement.