6-7
Monitor Display Options
Set Custom Defaults
When custom defaults are set up, they can be enabled or
disabled at any time through the Configure All or individual
Alarms/Targets Settings screen.
1 Touch the Settings button.
2 Touch Parameter Settings.
3 Touch Alarms / Targets.
4 Touch the Configure All button.
5 Touch the Set Custom Defaults button.
6 Touch the Parameter of interest.
7 Touch the value button for each target setting and enter
the desired value.
8 Continue steps 6 and 7 for each parameter. Touch the right
or left arrow at the bottom of the screen to display the next
or previous set of parameters.
9 Touch the Set all parameters according to: Indexed or
Non-Indexed.
10 When all desired parameters have been modified, touch
Confirm All.
Configure Targets and Alarms for One
Parameter
The Alarm Target Popup lets the user set up alarm and target
values for the selected parameter. The user can also enable or
disable the audible alarm and full alarm and target settings.
Adjust the target settings by using the numbered keypad or by
using the scroll buttons when a minor adjustment is needed.
1 Touch inside a globe to open the targets popup for that
parameter.
2 To disable the audible alarm for the parameter, touch the
Audible Alarm button at the top right of the popup.
3 To disable targets for the parameter, touch the Target
button at the top left of the popup.
4 Use the arrows to adjust the zone settings or touch the
value button to open a numeric keypad.
5 When the values are correct, touch the Enter button.
6 To cancel, touch the Cancel button.
Figure 6-9 Set Custom Default Alarms /
Targets
Figure 6-10 Set Alarms and Targets
WARNING
Physiological visual and audible physiological alarms
are activated only if the parameter is selected and
displayed on the screens as a key parameter (1-4
parameters). If a parameter is not selected and
displayed as a key parameter, the audible physiological
alarms are silenced.