Help and Troubleshooting
12-3
Pump-Unit Communication and Power
The Pump-Unit lights indicate the status of the system,
communication with the Monitor, Pump-Unit battery and AC
power. The Pump-Unit battery should not be removed or
tampered with. The internal battery will charge automatically
when the Pump-Unit is plugged into mains.
Figure 12-7 EV1000 Pump-Unit and Monitor LED Indicators
Pump-Unit alert
AC power status
Ethernet communication status
Monitor status
Battery power status
Table 12-1 Pump-Unit Communication and Power Lights
Condition Color Light Pattern Suggested Action
PUMP-UNIT ALERT LIGHT
No Alarm No light Solid OFF None
Alarm, severe error detected in the
Pump-Unit.
Red Flashing ON/OFF Turn the power off by disconnecting the main power
cord. The system will shut down in a few seconds.
Disconnect and reconnect all cables and power on
the system again using the monitor button.
If the problem persists, contact Edwards Technical
Support.

ETHERNET COMMUNICATION STATUS LIGHT
No Ethernet connection No light Solid OFF Connect the Pump-Unit to the Monitor using the
ethernet cable.
Ethernet communication active Green Solid ON None
Ethernet communication error Amber Flashing ON/OFF Check ethernet connections. If the problem persists,
contact Edwards Technical Support.

BATTERY POWER STATUS LIGHT
Battery charged Green Solid ON None
Battery charging Green Fading ON/OFF None
Battery power low Amber Solid ON Connect the main power cable to charge battery.