Methodology and Monitoring
7-5
4 Connect the ClearSight Finger Cuff to the Pressure
Controller.
Do not rotate the cuff after application. The hook located
on the outside of the finger cuff should be situated on the
back of the finger at all times.
5 If continuous monitoring is expected to last longer than 8
hours, or to increase patient comfort, repeat cuff sizing
and steps 1-4 to apply an additional cuff to a second finger
on the same hand.
Single Patient Use The ClearSight Finger Cuff is
designed for single patient use. Upon starting a measurement,
the Finger Cuff will expire after 72 hours for a single patient.
Double Cuff Application. The EV1000 NI allows two
ClearSight Finger Cuffs to be connected simultaneously to
alternate the measurement between two fingers. This feature
allows for continuous monitoring for durations of up to 72
hours and is required for measurements that take longer than
8 hours. This feature can also be used to increase patient
comfort. Accumulated monitoring on one finger is limited to
8 hours.
Enter Patient Data
1 Patient data can be entered upon initial startup of
the system or by touching the Settings button.
2 Touch Patient Data and enter the patient
demographics (see “Patient Data” on page 6-1).
3 Touch the Home button.
Zero and Apply Heart
Reference Sensor and Start
Monitoring
Before monitoring can be initiated, the Heart Reference
Sensor (HRS) must be zeroed and applied to the patient.
1 Connect the HRS to the Pressure Controller.
2 Touch the Clinical Actions button.
WARNING
Do not apply the ClearSight Finger Cuff or Pressure
Controller on injured skin as this may cause further
injury.
WARNING
Measurement on one finger in contradiction with the
instructions for use may affect patient comfort and/or
lead to minor injuries.
WARNING
To reduce the risk of skin irritation and tissue damage,
do not monitor longer than 8 hours continuously on a
single finger. To continue to monitor, apply the
ClearSight Finger Cuff to another finger or use two cuffs
to measure more than 8 hours.
WARNING
Do not use two ClearSight Finger Cuffs simultaneously
on the same finger.
CAUTION
Never bend a finger cuff to a flat shape, it will damage
the cuff and affect measurement accuracy
CAUTION
Excessive ambient light may interfere with ClearSight
Finger Cuff measurements.
CAUTION
The effectiveness of the ClearSight finger cuff has not
been established in pre-eclamptic patients.
Figure 7-5 Settings Screen
CAUTION
Make sure that the HRS is correctly applied so that it
can be leveled to the phlebostatic axis.