Help and Troubleshooting
ClearSight Warnings and Troubleshooting
Table 12-6 ClearSight Warnings
Message Possible Causes Suggested Actions
HRS Out of Range The HRS pressure offset exceeded limit
during the zeroing process.
Defective HRS.
Vertically align the two ends of HRS and re-zero.
Replace HRS.
If the problem persists, contact Edwards Technical
Support
HRS Zero Unsuccessful Prior to zero, no HRS movement detected.
During zero, HRS movement detected.
Disconnect and reconnect HRS.
Vertically align the two ends of HRS
Re-zero HRS.
Minimize movement of HRS ends during zeroing process.
Replace HRS and re-zero HRS.
If problem persists, contact Edwards Technical Support.
Unstable Arterial Pressure System detecting large variability in the
arterial pressure due to physiological or
artificial noise.
Ensure no external or artificial noise is interfering with
arterial pressure measurements.
Stabilize arterial pressure.
Large CO difference from
Reference
A large difference is detected between the
CO value and input Reference value.
Recalibrate the CO value.
Perform another Reference measurement.
Press Continue to proceed with user input Reference
value.
Calibration time has been
exceeded
The time between the completion of CO
averaging and user input Reference value
for calibration has exceeded system limits.
Perform a new CO calibration, ensuring a timely duration
between the completion of CO averaging, and the
acceptance of the Reference CO value.
Continuous Monitoring Has
Reached The 72 Hour Limit
Continuous monitoring has reached the 72
hour limit.
Perform monitoring on opposite hand.
Table 12-7 ClearSight Troubleshooting
Message / Question Possible Causes Suggested Actions
Connect Finger Cuff No Finger Cuff(s) detected.
Defective Finger Cuff(s) connected.
Connect Finger Cuff(s).
Replace Finger Cuff(s).
Finger Cuff #1 Approaching
Maximum Use Time
Finger Cuff #1 approaching maximum use
time.
Replace Finger Cuff #1 to ensure uninterrupted
measurement.
Finger Cuff #2 Approaching
Maximum Use Time
Finger Cuff #2 approaching maximum use
time.
Replace Finger Cuff #2 to ensure uninterrupted
measurement.
Connect HRS HRS connection not detected. Connect HRS.
Replace HRS.
Zero HRS HRS not zeroed. Ensure HRS is connected and Zero HRS to Start
Measurement.
Connect Pressure Controller Pressure Controller not connected.
Defective Pressure Controller connected.
Connect Pressure Controller.
Replace Pressure Controller.
If problem persists, contact Edwards Technical Support.
EV1000NI Use Not
Recommended For Patient Age
< 18 yrs
Non invasive BP measurement technology
not validated for patients under 18 years of
age.
Measurement with an alternate BP / Cardiac Output
technology recommended.
Pump-Unit Service Required Pump-Unit service time is approaching. Replace Pump-Unit.