EasyManua.ls Logo

Eickemeyer MAGIC 6000 - Page 162

Eickemeyer MAGIC 6000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12-22 Probes and Biopsy
4) Select a proper guiding block and push it into the groove above the angle block,
and clamp it tightly.
5) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
NGB-016
1. Put on the sterile transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the
transducer. Mount the bracket onto the transducer.
3. Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
4. Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
5. Screw the nut of the block to secure the block.
6. Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.

Table of Contents

Related product manuals