Acoustic Output Reporting Table (60601-2-37) B-3
Transducer Model: 75L53EA
Imaging Mode: B+M-mode
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1 cm
2
A
aprt
> 1
cm
2
Maximum Index Value 7.71E-01 3.03E-01 1.84E-01 4.46E-01
Associate
Acoustic
Parameter
p
r.a
1.92E+00
P 4.79E+00 4.79E+00 4.79E+00
Min of [P
Į
(z
s
), I
ta,Į
(z
s
)]
z
s
z
bp
z
b
1.50E+00
z at max.I
pi,Į
1.50E+00
d
eq
(z
b
) 3.51E-01
f
awf
6.35E+00 5.64E+00 5.64E+00
Dim of A
aprt
X
1.79E+01 1.86E+00
Y 3.83E+00 5.00E-01
Other
Informatio
n
t
d
2.02E-01
prr 2.00E+03
p
r
at max. I
pi
2.66E+00
d
eq
at max. I
pi
3.41E-01
I
pa,Į
at max. MI 2.03E+02
Focal
Length
FL
x
(cm) 1.00E+01 1.00E+01
FL
y
(cm) 2.00E+00 2.00E+00
Operating
Control
Conditions
Focus position
1.5cm 10cm 10cm 10cm
Display depth
12cm 12cm 12cm 12cm
Acoustic power
100% 100% 100% 100%
PRF 2000 2000 2000 2000
Working Frequency H10M 5M 5M H8M
Scan Line Mode Lower-density line and minimum scan angle

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer MAGIC 6000 and is the answer not in the manual?

Eickemeyer MAGIC 6000 Specifications

General IconGeneral
BrandEickemeyer
ModelMAGIC 6000
CategoryMedical Equipment
LanguageEnglish

Related product manuals