Loading...
Probes and Biopsy 12-23
CAUTION:
Ensure that all guide parts are seated properly prior to performing
a biopsy.
12.2.5 Biopsy Menu
Press <Biopsy> to show the biopsy menu.
Select biopsy bracket angle
If the needle-guided bracket supports more than one biopsy angle, you can select the
angle from the drop-down list.
Select Dot size
Move the cursor onto [Dot Size], press <Set> to select the dot size among Small, Medium
and Big.
Tips:
z The guide line is a dot line which consists of two kinds of dots, the distance between
two dots is depth dependent. Move the cursor onto the big dot, a numeral, which
represents the biopsy depth, is displayed.
z The biopsy guidezone adjusts along with image adjustments, such as image
inversion/rotations, zoom and depth changes.
z When the imaging depth and area are changed, the guide line will be adjusted.
Display or hide the guide line
Set [Biopsy Kit] on the menu to !Off" to hide the guide line.
Or, press <Biopsy> on the control panel to display or hide the guide line.
In B mode image menu, you can also display, hide or select the needle-guide bracket via
[Biopsy Kit] item.
12.2.6 Verify Biopsy Guide Line
WARNING:
1.
Prior to each biopsy procedure, be sure to verify the guide
line.
2.
If the needle is not consistent with the guide line, DO NOT
perform the biopsy procedure.

Table of Contents