3-2 System Preparation
3.3 Powering ON/ OFF
3.3.1 Powering ON
CAUTION:
To ensure safe and effective system operation, you must
perform daily maintenance and checks. If the system begins
to function improperly – immediately stop scanning. If the
system continues to function improperly – fully shut down
the system and contact Customer Service
Department or sales representative. If you use the system in a
persistent improperly functioning state – you may harm the
animal or damage the equipment.
Checking before Power-on
To check the system before turning on the system:
No. Check Item
<1>
The temperature, relative humidity and atmospheric pressure shall meet the
requirements of operating conditions. See "2.4.3 Environmental Conditions" for
details.
<2> There shall be no condensation.
<3>
There shall be no distortion, damage or dirt on the system and peripheral
devices.
If any dirt is found, cleaning shall be performed as defined in section !16.1.1
Cleaning the System#.
<4> There shall be no loose screws on the LCD or control panel.
<5>
There shall be no cable damage (e.g. power cord). Maintaining secure
connections to the system at all times.
<6>
The probes and probe cables shall be free of damage or stains.
See !12.1.5 Probes Cleaning and Disinfection" for details on probe cleaning
and disinfection.
<7>
No miscellaneous odds and ends are allowed to be attached or affixed to the
control panel.
<8>
Ensure that all connections are free from damage and remain clear of foreign
object blockages.
There shall be no obstacles around the system and its air vent.
<9> Probe cleaning and disinfection.
<10> The overall scanning environment and field must be clean.
Checking after it is turned on
Press the power button in the upper right corner on the control panel.
To check the system after the system is turned on:
No. Check Item
<1> There shall no unusual sounds or smells indicating possible overheating.
<2> There shall be no persistently displayed system error message.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer MAGIC 6000 and is the answer not in the manual?

Eickemeyer MAGIC 6000 Specifications

General IconGeneral
BrandEickemeyer
ModelMAGIC 6000
CategoryMedical Equipment
LanguageEnglish

Related product manuals