Acoustic Output Reporting Table (60601-2-37) B-9
Transducer Model: 65EB10EA
Imaging Mode: B+M-mode
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1 cm
2
A
aprt
> 1
cm
2
Maximum Index Value 5.25E-01 1.35E-01 9.77E-02 2.04E-01
Associate
Acoustic
Parameter
p
r.a
1.19E+00
P 1.63E+00 1.63E+00 1.63E+00
Min of [P
Į
(z
s
), I
ta,Į
(z
s
)]
z
s
z
bp
z
b
2.00E+00
z at max.I
pi,Į
1.86E+00
d
eq
(z
b
) 2.02E-01
f
awf
5.22E+00 5.13E+00 5.13E+00
Dim of A
aprt
X
1.03E+01 8.96E-01
Y 3.37E+00 5.00E-01
Other
Informatio
n
t
d
3.89E-01
prr 2.00E+03
p
r
at max. I
pi
1.65E+00
d
eq
at max. I
pi
1.94E-01
I
pa,Į
at max. MI 8.44E+01
Focal
Length
FL
x
(cm)
1.00E+01 1.00E+01
FL
y
(cm)
3.00E+00 3.00E+00
Operating
Control
Conditions
Focus position
2.5cm 10cm 10cm 1cm
Display depth
12cm 12cm 12cm 2.8cm
Acoustic power
100% 100% 100% 100%
PRF 2000 2000 2000 2000
Working Frequency H9M H9M H9M H9M
Scan Line Mode Lower-density line and minimum scan angle