B-18 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 35C50EA
Imaging Mode: B+M-mode
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1 cm
2
A
aprt
> 1
cm
2
Maximum Index Value 7.31E-01 7.29E-01 6.67E-01 1.48E+00
Associate
Acoustic
Parameter
p
r.a
1.16E+00
P 1.99E+01 1.99E+01 1.99E+01
Min of [P
Į
(z
s
), I
ta,Į
(z
s
)]
z
s
z
bp
z
b
2.00E+00
z at max.I
pi,Į
3.34E+00
d
eq
(z
b
) 4.34E-01
f
awf
2.58E+00 3.42E+00 3.60E+00
Dim of A
aprt
X
6.98E+00 7.80E-01
Y 1.26E+01 1.24E+00
Other
Informatio
n
t
d
8.35E-01
prr 2.00E+03
p
r
at max. I
pi
1.56E+00
d
eq
at max. I
pi
4.22E-01
I
pa,Į
at max. MI 5.75E+01
Focal
Length
FL
x
(cm)
4.00E+00 4.00E+00
FL
y
(cm) 8.00E+00 8.00E+00
Operating
Control
Conditions
Focus position
2cm 4cm 4cm 4cm
Display depth
6.5cm 6.5cm 6.5cm 6.5cm
Acoustic power
100% 100% 100% 100%
PRF 2000 2000 2000 2000
Working Frequency H6M 3.5M 3.5M H6M