5-4 Image Optimization
Operation To increase the gain compensation at an area of interest, move the TGC slider
to the right. To decrease the gain compensation at the corresponding area of
interest, move the TGC slider to the left.
About 1.5s after the adjustment is finished, the TGC curve disappears.
Effects Adjust the signal gain for the certain image area to get a balanced image.
Frequency
Description This function is used to select the operating frequency of the current probe,
the real-time value of which is displayed in the image parameter area in the
upper left corner of the screen, where !F" represents B mode frequency, and
!FH" represents harmonic frequency.
Operation Adjust it through [Frequency] on the image menu, wherein !H" means the
harmonic frequency.
Or adjust it in the image parameter area.
Values of frequency vary depending upon the probe types. Select the
frequency according to the detection depth and current tissue characteristics.
Effects The higher the frequency the better the near field resolution but the worse the
force of penetration.
Harmonic imaging enhances near field resolution and reduces low-frequency
and large amplitude noise, so as to improve small parts imaging.
A. power
Description Refers to the power of ultrasonic wave transmitted by the probe, the real-time
value of which is displayed in the image parameter area in the upper left corner
of the screen.
Operation Adjust through the [Acoustic Power] item in the image menu;
The adjusting range is 7%-100% in increments of 3%.
Effects Generally, increasing the acoustic power will increase the brightness and
contrast of the image as well as the force of penetration.
Impacts You should perform exams according to actual situation and follow the ALARA
Principle.
Focus
Description
Refers to adjustment of focus of the ultrasonic beams, symbols as #
# of
which will be displayed on the right of the image.
Operation Adjust the focus number through the [Focus Number] in the menu;
Adjust the focus position through the [Focus Position] in the menu;
In B Mode, focus number can be switched among 1-4.
Effects The area that is focused will be of a higher contrast and resolution to provide
a much clearer image.
Impacts The greater the number of focus, the slower the frame rate to image.
Imaging Display Adjustment
Description More information can be obtained without moving the probe or changing the
sampling position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer MAGIC 6000 and is the answer not in the manual?

Eickemeyer MAGIC 6000 Specifications

General IconGeneral
BrandEickemeyer
ModelMAGIC 6000
CategoryMedical Equipment
LanguageEnglish

Related product manuals