Operation Proline Promag W 400 EtherNet/IP
102 Endress+Hauser
11.5.2 Totalizer
The Totalizer submenu contains all the parameters needed to display the current
measured values for every totalizer.
Navigation
"Diagnostics" menu → Measured values → Totalizer
Totalizer
Totalizer value 1 to 3
Totalizer overflow 1 to 3
Parameter overview with brief description
Parameter Prerequsite Description User interface Factory setting
Totalizer value 1 to 3 In the Assign process variable
parameter in the Totalizer 1
to 3 submenu, one of the
following options is selected:
Volume flow
Mass flow
Displays the current totalizer
counter value.
Signed floating-point
number
0 l
Totalizer overflow 1 to 3 In the Assign process variable
parameter in the Totalizer 1
to 3 submenu, one of the
following options is selected:
Volume flow
Mass flow
Displays the current totalizer
overflow.
Integer with sign 0
11.6 Adapting the measuring device to the process
conditions
The following are available for this purpose:
Basic settings using the Setup menu
Advanced settings using the Advanced setup submenu
11.7 Performing a totalizer reset
Function scope of "Control Totalizer " parameter
Options Description
Totalize The totalizer is started.
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.
Preset + totalize The totalizer is set to the defined start value in Preset value parameterand the
totaling process is restarted.