Commissioning Proline Promag W 400 EtherNet/IP
94 Endress+Hauser
ECC cleaning cycle
ECC Polarity
Parameter overview with brief description
Parameter Description Selection / User entry / User
interface
Factory setting
Electrode cleaning circuit Enable the cyclic electrode cleaning circuit. • Off
• On
Off
ECC duration Enter the duration of electrode cleaning in
seconds.
0.01 to 30 s 2 s
ECC recovery time Define recovery time after electrode
cleaning. During this time the current output
values will be held at last valid value.
1 to 3.0
+38
s 5 s
ECC cleaning cycle Enter the pause duration between electrode
cleaning cycles.
0.5 to 168 h 40 min
ECC Polarity Select the polarity of the electrode cleaning
circuit.
• Positive
• Negative
Depends on the electrode
material
10.7.5 Administration configuration
The Administration submenu contains administrative parameters.
Navigation
"Setup" menu → Advanced setup → Administration
‣
Administration
Define access code
Device reset
Parameter overview with brief description
Parameter Description User entry / Selection Factory setting
Define access code Restrict write-access to parameters to
protect the configuration of the device
against unintentional changes via the local
display.
0 to 9 999 0
Device reset Restart or reset device manually. • Cancel
• To delivery settings
• Restart device
Cancel
10.8 Configuration management
After commissioning, you can save the current device configuration, copy it to another
measuring point or restore the previous device configuration.
You can do so using the Configuration management parameter and the related options
found in the Configuration backup display submenu.