Proline Promag W 400 EtherNet/IP Diagnostics and troubleshooting
Endress+Hauser 111
12.4 Diagnostic information in the Web browser
12.4.1 Diagnostic options
Any faults detected by the measuring device are displayed in the Web browser on the
home page once the user has logged on.
1
3
2
A0017759-EN
1
2
3
Status area with status signal
Diagnostic information →  109
Remedial measures with Service ID
Furthermore, diagnostic events that have occurred can be viewed in the Diagnostics
menu:
Via parameters →  116
Via submenus →  117
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Symbol Meaning
A0017271
Failure
A device error has occurred. The measured value is no longer valid.
A0017278
Function check
The device is in service mode (e.g. during a simulation).
A0017277
Out of specification
The device is operated:
Outside its technical specification limits (e.g. outside the process temperature range)
A0017276
Maintenance required
Maintenance is required. The measured value is still valid.
The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107.
12.4.2 Calling up remedy information
Remedy information is provided for every diagnostic event to ensure that problems can be
rectified quickly. These measures are displayed in red along with the diagnostic event and
the related diagnostic information.